MODEL 924 - INSERTION MOUNTING

The insertion assembly allows pH measurements to be made in pressurized tanks and main
lines without the need for system shutdown for electrode maintenance. Its adjustable insertion
depth allows it to be positioned in a turbulent flow region so as to operate in the self-cleaning 
mode. In abrasive applications, the electrode is usually positioned even with the pipe wall. It is
well suited for measurements in liquids with high suspended solids and for measurements in 
viscous liquids.

A ball valve, packing gland arrangement allows the electrode to be installed and removed
from the pressurized system without shutting the system down. The special electrode used in
the insertion assembly allows it to be mounted in any position; for example, through the
bottom of a tank with its measuring surface facing upwards.











Go to Main Page    Goto pH Sensor Selector